Ionization Test Set
Perfect Kit for Testing and Balancing Ionizers
- Ionization Test Kit for Measuring Offset Voltage and Discharge Times
- Includes a Field Meter, Charge Plate Monitor, ±1000 Volt Charger and Decay Timer
- Balance Ionizers per ANSI/ESD SP3.3 Periodic Verification
- Measure Ionizer Decay time as per ANSI/ESD S3.1 and ANSI/ESD SP3.3.
- Measure Body Voltage Generation and Decay
- Evaluate Ionizers and Packaging Materials
- Includes a fixture to hold the PFM-711B, CPM-720B and PDT-740B as one
- Molded Carying Case with Foam Insert IncludedOverview
What's Included
- PFM-711B Electrostatic Field Meter
- PCS-730B Electrostatic Charger
- PCS-730BW 1kV Charging Rod
- CPM-720B Charge Plate Monitor Assembly
- PDT-740B Static Decay Timer
- 700AC Analog Cable
- 9 Volt Alkaline Batteries (3)
- PIK-112C Molded Carrying Case
Overview
Measures Offset Voltage and Charge Decay
The PFK-105 Ionization Test Kit is a convenient kit that includes the PFM-711B Field Meter, CPM-720B Charge Plate Monitor, the PCS-730B ±1 kV Charger and PDT-740B Static Decay Timer for measuring the perofrmance of ionizers in the area of use. Included with the kit a lightweight fixture allowing to combine the Field Meter, Charge Plate and Decay Timer into one hand-held unit, making it easier to measure the offset voltage and discharge times of your ionizers.
Meter for Testing of Electrostatic Fields
The hand-held PFM-711B Fieldmeter is an instrument used to perform static surveys and polarity on objects with accuracy of ±5% at ambient conditions (23ºC and 30%Rh). This stabilized chopper can be used in environments with a presence of ionization without drifting. The Hold button allows the field meter to hold the measurement for easy recording. It also features automatic polarity indication, low battery indicator, high and low ranges, distance ranging lights for accurate and repeatable measurements, a Zero control button and a ground snap. The PFM-711B includes two ranges of operation:
- 0 to 1,999V in 1 Volt increments with <±5% error
- 0 to 20,000V in 10 Volt increments with <±5% error up to 5kV
The PFM-711B can be used with a X-Y plotter, any analog data-recording devices, or the PGA-710B Autoanalysis System. The analog output provides a 1 volt output corresponding to 10kV, which is independent of the selected meter range and indication.
Charge Plate Monitor allowing you to Measure Ionizers
The CPM-720B Charge Plate Monitor is a conductive plate that attaches to the PFM-711B Field Meter. When attached, the system will allow to balance ionizers per ANSI/ESD SP3.3, voltage or charge generation and voltage suppression analysis.
Apply a Charge of ±1,000 volts
The PCS-730B ±1 kV Electrostatic Charger is used to charge the CPM-720B Charge Plate Monitor to ±1,000 volts to measure ionizer decay time as required by both ANSI/ESD S3.1 and ANSI/ESD SP3.3.
The PFM-711B Field Meter, the CPM-720B Charge Plate Monitor and the PCS-730B Charging Source are match calibrated to work together for accurate and efficient auditing.
Measuring Decay Time
The PDT-740B Static Decay Timer is typical of Prostat's innovative technology and support of professional ESD control managers and auditors.
This unique accessory is designed to measure the decay time of ionizers from ±1,000 volts to less than 100 volts in accordance with ESD Association's Ionization Standard ANSI/ESD S3.1, and measure material decay time to less than 50 volts and less than 10 volts.
A Helping Hand for Ionizer Analysis!
The PFK-100H fixture is designed to easily hold your PFM-711B Field Meter, your CPM-720B Charge Plate Monitor and your PDT-740B Static Decay Timer. Lightweight, this instrument holder will facilitate balancing your ionizer, by holding all instruments in 1 hand.
Complies to ANSI/ESD Standards
The PFK-105 will assess the performance specifications of ionizers in accordance with the following ESD Association documents:
- ESD TR53
- ANSI/ESD S3.1
- ANSI/ESD SP3.3
- ANSI/ESD S20.20
Technical Specifications
PFM-711B Electrostatic Field Meter Specifications
- Performance
- Overall Range:
- Low Range: 0 V to ±1999 V
- High Range: 0 V to ±19.99 kV
- Accuracy:
- Voltage Monitor Output: Better than ±5% of reading, ±10 mV
- Voltage Display: Better than ±5% of reading, ±2 counts
- Stability: ±10 counts
- Overall Range:
- Voltage Monitor
- Output: ±20.0 kV meter reading equals ±2.0 Volt output
- Ratio:
- Low Range: 0 to 2,000 volts is equal to ±0.2 volts output
- High Range: 0 to 20,000 volts is equal to ±2 volts output
- Front Panel
- Display:
- Number of Characters: 3.5 digits
- Character Format: 7-Segment
- Character Size: 12.70mm H x 7.62mm W
- Automatic polarity, LOW BATTERY indicator
- Range:
- V / Inch: 0 to ±1999 V / inch
- kV / Inch: 0 to ±19.99 kV / inch
- Display Resolution
- Low Range: 1 Volt increment
- High Range: 10 Volts incrementsh
- Sampling Rate: Display readings are updated every 0.5 seconds. Analog readings are taken every 100mS, or 10 times per second.
- Low Battery: Low battery indicator comes on at 7.4 Volts
- Display:
- Power
- Inout: 9 VDC alkaline battery (included)
- Battery Life: Approx. 20 hours (of continuous use)
- Grounding
- Case: Circuit grounded through conductive case
- Ground Snap: 10 mm snap fastener mounted on back of case
- Measurement Distance: 1" ± 1/64" (25mm ± 0.5mm)LED distance indicator; aligned targets indicate 1 inch (25 mm)
- Controls
- Keypad: A custom membrane controls the different functions
- ON/OFF: Short press to turn the instrument ON, Long press to turn the instrument OFF
- Range: Selects between V/Inch (0-2000V) and kV/Inch (0-20kV)
- Zero: Adjusts the displayed value to Zero (0)
- Hold: Captures and holds a stable measurement
- Output
- Analog Output: 2.5 mm jack (3/32") audio jack, 2 Volts output corresponds to 20 kV
- Response Time: Less than 180ms (90-10%)
- Sensor Noise Signal (Typical)1:
- Maximum Voltage: +3 volts
- Minimum Voltage: -5 volts
- Environment
- Operating Temperature: 10°C to 30°C (50°F to 80°F)
- Operating Relative Humidity: 0% - 80%, non-condensing
- Operating Altitude: 2000 m
- Physical Specifications
- Size (with ground snap attached): 27.2 mm H x 61.5 mm W x 105.9 mm L (1.07" H x 2.41" W x 4.17" L)
- Weight: 142 g (5.0 oz.) with battery
- Warranty
- 2 years on circuit board, display and membrane
- 1 year on sensor2
1Typical specifications are not guaranteed.
2Warranty on the sensor limited to a defective sensor that was not dropped or used for measuring a source greater than 20kV.
CPM-720B Charge Plate Monitor Assebly Specifications
- Performance
- Tolerance: Matches the PFM-711B Field Meter within 5% when calibrated together
- Capacitance: 19 Picofarads (pF) ±1 pF when mounted to the PFM-711B
- Range: 0 to ±5 kV ±2%
- Grounding
- Case: Through conductive case and snap fastener of PFM-711B Field Meter
- Environment
- Operating Temperature: 10°C to 30°C (50°F to 80°F)
- Operating Relative Humidity: 0% - 70%, non-condensing
- Operating Altitude: 2000 m
- Physical Specifications
- Without the Banana Input Jack: 8.25 cm L x 8.25 cm W x 1.75 cm H (3.25" L x 3.25" W x 0.69" H)
- With the Banana Input Jack: 8.25 cm L x 8.25 cm W x 2.29 cm H (3.25" L x 3.25" W x 0.90" H)
- Weight: 187 g (6.6 oz.)
PCS-730B Electrostatic Charger Specifications
- Performance
- Output: ±1,250 ±5% volts DC, Current limited to less than 1 microamp
- Output Connector: Female banana jack provided with mating contact rod
- Charging Speed: Charges from zero to over ±1000 volts in 3 seconds at ambient conditions
- Power
- Inout: 9 VDC alkaline battery (included)
- Battery Life: Minimum of 28 hours (of continuous use), Approximately 20,000 charges of either polarity
- Grounding: 10mm female snap fastener on rear of case
- Environment
- Operating Temperature: 10°C to 30°C (50°F to 80°F)
- Operating Relative Humidity: 0% - 80%, non-condensing
- Operating Altitude: 2000 m
- Physical Specifications
- Electrostatic Charger: 2.85 cm H x 6.12 cm W x 12.65 cm L (1.12" H x 2.41" W x 4.98" L)
- Charging Rod:
- Length 4.45cm (1.75")
- Length with banana plug: 8.18 cm (3.22")
- Diameter: 0.154"
- Weight: 147 g (5.2 oz.) with battery
- Warranty: 2 Year Limited Warranty
PDT-740B Static Decay Timer Specifications
- Controls: OFF-ON Cut Off Voltage, 10, 50 and 100 Volts (Selects Voltage to stop timing)
- Indicators:
- Green LED: Decay Cycle Ready > ±1kV, CPM voltage > ±1,000 Volts
- Red LED: Timing in Progress, CPM voltage decay timing activated
- LOBAT: Displayed when battery voltage < 6.0 Volts
- Display: Liquid Crystal Display (LCD) displays elapsed decay time from ±1,000 Volts to selected Cut Off voltage in seconds and tenths of a second (00.0)
- Polarity: Automatically senses Positive & Negative voltage references
- Zero & Reset: Automatically Resets Timer and Display to 00.0 when Charge Plate Monitor voltage > ±1kV. Reset indicated when Green LED Decay Cycle Ready LED is ON
- Response: Approximately 200 milliseconds in combination with Field meter and Charge Plate Monitor
- Timing Range: From 0.2 to 159.9 seconds
- Timing Setpoints:
- Reset when voltage input 100mv ±2 mv (1,000 Field Meter Volts ±20 V)
- Cut Off Voltage @ 100 V: 10mv ±0.2 mv (100 Field Meter Volts ±2 V)
- Cut Off Voltage @ 50 V: 5mv ±0.2 mv (50 Field Meter Volts ±2 V)
- Cut Off Voltage @ 10 V: 1mv ±0.2 mv (10 Field Meter Volts ±2 V)
- Accuracy: Typically < ±5%
- Power: 9 VDC alkaline battery (included)
- Battery Life: > 40 hours typical
- Dimensions: 11.4 cm L x 7 cm W x 2.5 cm H (4.50" L x 2.75" W x 1.0" H)
- Weight: 184 g (6.5 oz) with battery
- Warranty: 1 Year Limited Warranty
PFK-100H Instrument Holder Specifications
- Construction: Aluminum
- Dimensions: 20.96 cm x 7.62 cm x 2.54 cm (8.25" x 3.0" x 1.0")
- Weight: 153 g (5.4 oz)
SKU | I339912 |
---|---|
Mfg. Part Number | PFK-105 |
Brand | Prostat |