Programmed one touch buttons for common sterilization process; Manual, Melt -> Warm, Sterilize -> Warm, Sterilize/Liquid, Sterilize/Normal.
In addition to default sterilization programs, custom-programming can be saved for repeat uses.
To minimize cooling time, a cooling fan is provided for all models as standard equipment.
Two sensor ports are provided as standard on the chamber.
The drain bottle is located in front for easy access and drain water level can be monitored without opening the cabinet door.
A large-diameter drain pipe is provided for easier maintenance.
Safety Features: Steam cup, interlocking lid, self-diagnostic, alarm warning in the absence of drain bottle, circuit breaker, during any malfunction, the autoclave will automatically shut down the alarm buzzer sounding and error message displayed
System: Automatic high-pressure steam sterilization system for laboratory use
Range of working temperatures: 5-135°C: 45-60°C for preheating, 40-60°C for keep warm, 65-100°C for melting, 105-135°C for sterilization
Max. working pressure: 0.255MPa
Ambient temperature: 5 - 35°C
Lid: Manual upward opening with an interlock for safety
Heater: 100V, 1.9kW (950W 2 units)
Exhaust valve: An exhaust valve and a slow release valve
Connection ports for optional accessories: 1/4" female thread for, sample temp. sensor, chamber temp. sensor, and pressure sensor (branching from the solenoid valve tubing) Total: 3 ports
Cooling fan: Axial fan motor
Temperature setting, control and display: PID control by microprocessor, digital setting (by UP/DOWN keys) and display
Timer and its resolution: 0 or 1 minute to 99 hours 59 minutes in increments/decrements of 1 minute
Operation mode: Instrument sterilization course, liquid sterilization course, sterilization and keep/warm course, melting and keep/warm course, customer-programmed course
Other features: Key locking, presetting, saving, preheating, forced cooling, sample temperature sensor (option), memory locking, saving the history of up to 20 errors, indication of accumulated working hours, indication of present time, on-off beeping setting
Safety measures: Sensor malfunctioning, SSR short-circuit, broken heater wire, prevention of idle heating (liquid expansion type), alarm against the absence of a drain bottle, failure in locking the cover, memory malfunctioning, pressure relief valve (0.255 MPa)
Pressure vessel standard: Small-sized pressure vessel (notification of installation is not required)
External dimensions (inches) WxDxH: 18.1 x 23.2 x 41.6
Effective dimensions of chamber (in): 11.8 (inner diameter) x 26.1 (depth)
Effective Capacity of chamber: 47L
Weight: 85kg
Power supply: 100V-120V AC
19.5A-23.5A
Accessories: Rack x 3, OSM-70 (241088)(dia 10.4 x depth 8.0 in), Vapor cup, Drain bottle, Drain board, Filter, Instruction manual